Abstract

Force plates for human movement analysis provide accurate measurements when mounted rigidly on an inertial reference frame. Large measurement errors occur, however, when the force plate is accelerated, or tilted relative to gravity. This prohibits the use of force plates in human perturbation studies with controlled surface movements, or in conditions where the foundation is moving or not sufficiently rigid. Here we present a linear model to predict the inertial and gravitational artifacts using accelerometer signals. The model is first calibrated with data collected from random movements of the unloaded system and then used to compensate for the errors in another trial. The method was tested experimentally on an instrumented force treadmill capable of dynamic mediolateral translation and sagittal pitch. The compensation was evaluated in five experimental conditions, including platform motions induced by actuators, by motor vibration, and by human ground reaction forces. In the test that included all sources of platform motion, the root-mean-square (RMS) errors were 39.0 N and 15.3 N m in force and moment, before compensation, and 1.6 N and 1.1 N m, after compensation. A sensitivity analysis was performed to determine the effect on estimating joint moments during human gait. Joint moment errors in hip, knee, and ankle were initially 53.80 N m, 32.69 N m, and 19.10 N m, and reduced to 1.67 N m, 1.37 N m, and 1.13 N m with our method. It was concluded that the compensation method can reduce the inertial and gravitational artifacts to an acceptable level for human gait analysis.

Introduction

Force plates and instrumented treadmills are commonly used to measure ground reaction forces (GRF) for clinical movement analysis, sports performance, or research on human movement. The combination of motion capture and force plate data allow for the calculation of joint moments through inverse dynamic analysis. Recently, instrumented treadmills have become equipped with actuators to translate and rotate the walking surface, for either virtual reality applications or for testing human response to perturbations. Acceleration of the force plate creates large inertial artifacts in the GRF measurement, because a large moving mass is located between the force of interest (foot/ground interface) and the load cells. Additionally, when the frame is tilted, its gravitational mass starts contributing to the horizontal GRF signals. The problem is especially severe in an instrumented treadmill, where the moving mass includes the treadmill frame, motor, and belts. These large errors in GRF data make it impossible to perform standard inverse dynamics in these conditions because the joint moment calculations are based on inaccurate force measurements. The same problem also occurs when the force plate foundation is not sufficiently rigid, or in a moving vehicle in order to study the biomechanics of driving.

Although the problem is noteworthy in human movement analysis, it exists in any load measurement system where the force of interest and the load cells are separated by a moving mass. This includes certain applications in high-speed material testing and in force-controlled robots (Hessling, 2009, Dixon, 1990). The work presented in this paper will be applicable in those fields as well.

In principle, the inertial and gravitational forces can be estimated and compensated using rigid body dynamics. This requires knowledge of the mass, inertia matrix, acceleration, angular acceleration, angular velocity, and orientation of the frame. This has been successfully done for one-dimensional linear motion such as in materials testing or a sliding force plate (Hessling, 2009, Dixon, 1990, Pagnacco et al., 2000, Yang and Pai, 2006). While it is straightforward to extend this approach into a six degree of freedom (DOF) load measurement, it becomes impractical due to the requirement to estimate full 3D motion relative to an inertial reference frame, and the use of nonlinear models (Berme and Guler, 2012a, Berme and Guler, 2012b, Hou et al., 2009, van den Bogert et al., 1996). Furthermore, mass and inertial properties of the frame must be known (Preuss and Fung, 2004). Some of the existing methods neglect the effect of rotation and are limited to compensating for errors due to vibrations within the building or floor (Boschetti et al., 2013).

In this paper, we introduce a simple linear, accelerometer-based compensation method for a fully general inertial and gravitational compensation of force plate data. The linear model is based on the principle that an accelerometer directly measures the inertial and gravitational force on its internal test mass. With a sufficient number of accelerometers, attached at different locations, the total inertial and gravitational artifact of all mass elements in the moving frame will be a linear combination of accelerometer signals (Zappa et al., 2001). The method will be presented and evaluated on an instrumented treadmill in various experimental conditions.

Section snippets

Compensation method

In a 6-DOF load measurement, three-dimensional force and moment are the variables of interest. The compensation model assumes that the effect of gravity and inertia on each of the six load signals, when expressed in the local reference frame of the sensor, is a static linear function of N accelerometer signals:FxFyFzMxMyMz=Ca1a2aN1,where C is a 6×(N+1) matrix of model coefficients. Note that the model includes a constant term for each load variable (the last column of C) which will be used by

Results

Fig. 3 illustrates the performance of the compensation method on the first four seconds of Trial 3 which contains artifacts due to actuated random motions, human ground reaction forces, and motor vibration. Before compensation, there are large artifacts, especially in Fx,Fz,Mx, and My. The artifacts in Fz and My are mostly gravitational and reflect low-frequency changes in pitch angle. The inertial contributions have higher frequencies, reflecting the sensitivity to linear and angular

Discussion

Rotating and translating a platform equipped with load cells will introduce inertial artifacts in the force measurements. We developed a linear model to estimate these errors from accelerometers so that these errors can be compensated. The method was evaluated on the test case of human gait analysis on an instrumented treadmill undergoing movement. It was found that the errors were initially too large to allow a useful inverse dynamic gait analysis. The compensation method reduced the errors to

Conclusion

The proposed method is capable of reducing inertial and gravitational artifacts in data obtained from a moving instrumented treadmill. For gait applications, the remaining error is small enough to allow accurate joint torque calculations through inverse dynamic analysis.

Conflict of Interest Statement

There are no conflicts of interest for any of the authors regarding the research reported in this manuscript.

Acknowledgment

This research was supported by the National Science Foundation under Grant No. 1344954 and by the Ohio Department of Development, Third Frontier Commission. The authors thank Co van Vliet for mounting and wiring the accelerometers.

References (18)

There are more references available in the full text version of this article.

Cited by (5)

  • Development of an integrated biomechanics informatics system with knowledge discovery and decision support tools for research of injury prevention and performance enhancement

    2022, Computers in Biology and Medicine
    Citation Excerpt :

    In studies involving foot strike and associated forces, the forces collected from a force plate on the ground versus those collected from different instrumented treadmills will have inherent differences and require pre-processing to standardize the data [10]. For example, force plates that are embedded within platforms that are moving or tilted will contain artifacts [11,12]. There will also be differences in the velocities of the center of gravity between overground and treadmill running [13].

  • A method to estimate inertial properties and force plate inertial components for instrumented platforms

    2019, Medical Engineering and Physics
    Citation Excerpt :

    A comparison between the predicted and acquired force plate signals validated the use of motion capture and inverse dynamics to reliably reduce FPIC from force plate data collected under dynamic conditions. Other potential methods to remove FPIC use accelerometers instead of motion capture [17,18]. Given that the CAREN is already equipped with a motion capture system, the approach outlined by Preuss and Fung [14] offers the most suitable option to estimate and remove FPIC.

View full text